- Augment environmental control of chambers to extend experimental capability
- Combine temperature control and relative humidity control within the same stage and the same experiment
FEATURES
• INSTEC's RHC01 Relative Humidity Controller is designed to augment environmental control of chambers to extend experimental capability
• Combine temperature control and relative humidity control within the same stage and the same experience
• Pairs seamlessly with the TS102GXY, HCS621GXY, or HS1200GXY, and many other stages with sealed chambers
• Test and characterize dependence of chemical properties, surface features, sample behavior, and device function on relative humidity
• Precision control of relative humidity from 4% to 91% within sealed chambers with volumes of up to 2000cc
• Recycling desiccant system dries ambient air
• Continuous operation possible for days, weeks, and months at a time with easy refillable water bottle
• Can be used as standalone humidity generator and controller or with INSTECAPP software to record and export sensor data
• Sensor Type: combined temperature and humidity digital capacitance sensor
• One sensor monitors the humidity of the air delivered from the controller
• Another sensor monitors the humidity of the air in the chamber and connect via a feedback loop to the controller
• Does not require an expensive dry air supply
• Optional inert gas regulator to minimize air and oxygen supplied to the chamber
• Interactive touchscreen display to control all features
SPECIFICATIONS
Humidity range |
4% to 91% |
Humidity sensor accuracy |
±1.8% |
Humidity resolution |
±0.5% |
Humidity stability |
±0.5% |
Maximum ramp rate |
+7% per minute |
Humidity sensor delay |
8 seconds |
Maximum controllable volume |
2000 cm3 |
Maximum gas pressure |
1.5 bar |
Temperature range |
20°C to 60°C |
Temperature accuracy |
±0.5°C at 25°C |
Temperature resolution |
0.1°C |
Input power supply |
110 V or 230V depending on region |
Dimensions |