High temperature microscopy & spectroscopy plate
 - 
	
30°C ~ 1200°C
 - 
	
Φ12 mm sample area
 - 
	
Gas-tight chamber *Vacuum-tight model available (HP1200V)
 
DESCRIPTION
Instec의 HP1200 Hot Gas-Tight 스테이지는 샘플을 1200 °C까지 가열할 수 있으며 세라믹, 야금, 지질학 및 고온 재료 연구에 이상적입니다. Electrical feedthrough 및 맞춤형 샘플 영역 (sample area)이 가능합니다. HP1200G의 gas-tight 샘플 챔버는 고온에서 샘플 산화를 방지하는 데 도움이 됩니다.
THERMAL FEATURES
| 
			 Temperature Control  | 
			
			 mK2000 with programmable precision switching PID method  | 
		
|---|---|
| 
			 Thermal Block  | 
			
			 Ceramic  | 
		
| 
			 Sample Thermal Cover  | 
			
			 Removable Inner sample cover with additional window  | 
		
| 
			 Temperature Minimum  | 
			
			 30°C (active cooling not available)  | 
		
| 
			 Temperature Maximum  | 
			
			 1200°C  | 
		
| 
			 Temperature Sensor  | 
			
			 S Type Thermocouple  | 
		
| 
			 Maximum Heating Rate  | 
			
			 +200°C per minute at 100°C  | 
		
| 
			 Minimum Heating and Cooling Rate  | 
			
			 ±0.5°C per minute  | 
		
| 
			 Temperature Resolution  | 
			
			 0.1°C  | 
		
| 
			 Temperature Stability  | 
			
			 ±1°C  | 
		
| 
			 Power supply  | 
			
			 Universal power input  | 
		
| 
			 Software  | 
			
			 Windows compatible software to record and export temperature-time data; remotely control system via VCP connection (Virtual Comm Port)  | 
		
OPTICAL FEATURES
| 
			 Optical access  | 
			
			 Reflection capability  | 
		
|---|---|
| 
			 Optical windows  | 
			
			 Removable and exchangeable windows permit full-spectrum transparency  | 
		
| 
			 Minimum Objective Working Distance  | 
			
			 9.0 mm  | 
		
| 
			 Top Aperture  | 
			
			 38 mm diameter  | 
		
| 
			 Top Viewing Angle  | 
			
			 ±24.5° from normal  | 
		
STRUCTURAL FEATURES
| 
			 Sample Area  | 
			
			 Φ12 mm x 5 mm  | 
		
|---|---|
| 
			 Atmosphere Control  | 
			
			 Gas purging capability to control humidity, condensation, and oxidation  | 
		
| 
			 Frame Cooling  | 
			
			 Integrated frame cooling with optional chiller system  | 
		
| 
			 Window and objective protection  | 
			
			 Cooling by purge air  | 
		
| 
			 Mounting  | 
			
			 Horizontal and vertical mounting  | 
		
| 
			 Frame Dimensions  | 
			
			 200 mm x 120 mm x 30 mm  | 
		
| 
			 Weight  | 
			
			 1350 g  | 
		
Instec's HP1200 Hot Gas-Tight stage can heat samples to 1200°C and are ideal for the study of ceramics, metallurgy, geology and high temperature materials. Electrical feedthroughs and custom sample areas are also available. The HP1200G's gas-tight sample chamber helps to prevent sample oxidation at high temperatures.