Highlights
- Dual furnaces structure, the system consists of two parts, one for moisture and ash analysis, other for volatile matter analysis.
- For moisture and ash analysis, 19 samples can be loaded once + 1 blank sample position.
- For volatile matter analysis, vertical furnace structure is adopted with 26 positions. Upper part for high temperature volatile matter test while lower part for low temperature cooling and weighing, makes the test result more accurate and test efficiency greatly improved.
- Flexible software design, test procedure, analysis time, temperature etc. can be customized.
- Unique sample weighing mode, single sample weighing and batch sample weighing are both available, makes the operation convenient.
- After finishing the moisture and ash test, furnace cover will be opened automatically for furnace cooling down, thus to shorten the waiting time for next batch of test, makes the operation safe and efficient.
- LED displays on both furnaces to show real-time balance weight data, it makes the sample weighing more convenient.
- Automatic temperature calibration function is available.
Specification
Description |
SDTGA6000 (include SDTGA6000A and SDTGA6000V) |
Test index |
Moisture, ash and volatile matter |
Continuous test in batches |
Yes |
Standards |
ISO 11722, ISO 1171, ISO 562, ASTM-D 5142-2009 |
Precision |
0.02% RSD (1σ, 1g sample) |
Furnace |
Dual furnaces, one for moisture and ash, one for volatile matter |
Furnace temperature |
Room temperature ~ 1000℃ |
Temperature control precision |
± 2℃ |
Temperature resolution |
0.1℃ |
Reproducibility |
Ash: 0.30% (<15.00%); 0.50% (15.00~30.00%); 0.70% (>30.00%) |
Cooling time required for batches of test |
No |
Furnace Cooling (forced air/N2) |
950 to 500°C <30 min |
Balance Resolution |
0.0001g |
Sample introduction and taking |
Automatic |
Max. Samples |
19 samples + 1 blank sample position (Moisture & Ash) |
26 samples + 1 blank sample position (Volatile matter) |
|
Quantity of sample / test time: |
Full indexes for 19 samples< 120 min (Quick method) |
Sample weight |
0.5 up to 5g |
Max. power |
4 kW for moisture/ash testing; |
Interface: |
RS232 |
Furnace Heating |
Resistance Wound |
Temperature Calibration |
Standard thermocouple |
Temp. control precision |
± 2℃ for moisture |
Gas Flow Rate |
Moisture:Nitrogen or Air 2~4(Furnace Volume /min) |
Power requirement |
220VAC ±10%,50 / 60Hz |
Gas requirement |
Nitrogen / Oxygen / Compressed Air |
Gas Pressure |
Air 5 to 6 bar (75 to 90 psi) |
Minimum Gas Purity |
Air/Nitrogen/Oxygen 99.5%; dry, oil-free |
Sample weighing |
Under normal temp. |
Test Temperature |
Can be customized |
Humidity |
<80% |
Dimensions |
Moisture & ash analyzer: 550 x 590 x530mm |
Weight |
Moisture & Ash analyzer: 70kg |
Operational Control |
Windows7/ Windows10 on external PC |
Advance Calculation |
Yes |
Analysis Report |
Yes |
Statistics Report |
Yes |
System Monitor |
Yes |