Highlights
High precision & accuracy
- Two internal balances away from high temperature and heat radiation, no drift occurs.
- With automatic calibration function, each weighing is accurate and reliable.
- Simulate test environment and condition of drying oven and muffle furnace, exact in accordance with standard.
- The unique automatic cleaning function of low temperature furnace completely eliminates the influence of dust on result.
High efficiency
- Dual furnaces design, low-temperature furnace for moisture test and high- temperature furnace for ash and volatile matter test.
- Analysis for moisture, ash and volatile matter can be conducted simultaneously and consecutively.
- No blank sample during test. Full index of single batch samples (18 samples) can be finished in 150 min.
Patented technology
- Tailored volatile matter crucible with unique auto-sample loading mechanism avoided the operation of replacement or removal of the crucible cover under high temperature.
- Unique crucible clamp device avoid upset and drop of crucible.
- Unique crucible detection mechanism avoid mistake and missing of placing crucible.
- Unique sample tray positioning mechanism and drive mechanism ensure stable and reliable operation.
Human based design, Environment friendly and energy saving
- Sample tray locates inside, real time data transmission. Weighing sample at room temperature, away from heat radiation.
- Special crucible collector discards residual automatically. No need to clamp crucible at high temperature. Easy and safe operation.
- Equipped with dust collector, environment friendly and clean.
- Compressed air can meet the requirement of gas supply, no oxygen and nitrogen required.
Intelligent control
- Easy to use Windows-based operating software, after finishing the sample weighing, the operator can be free for other tasks.
- Several proximate analyzers can be controlled by a single PC with Ethernet communication technology.
- Strong data processing ability, statistic report and print function, connection with network and balance are available by standard interface RS232.
Specification
Description
|
SDTGA8000
|
---|---|
Test index
|
Moisture, ash and volatile matter
|
Continuous test in batches
|
Yes
|
Standards
|
ISO 11722, ISO 1171, ISO 562, ASTM-D 5142-2009
|
Precision
|
0.02% RSD (1σ, 1g sample)
|
Reproducibility
|
Ash: 0.30% (<15.00%); 0.50% (15.00~30.00%); 0.70% (>30.00%)
Volatile matter: 0.50% (<20.00%); 1.00% (20.00~40.00%); 1.50% (>40.00%) |
Cooling time required for batches of test
|
No
|
Furnace Cooling (forced air/N2)
|
950 to 500°C <30 min
|
Balance Resolution
|
0.0001 g
|
Quantity of furnace
|
2
|
Sample introduction and taking
|
Automatic
|
Quantity of sample/ Analysis time
|
16 samples / 60 min~90 min (Moisture);16 samples/ 230 min (America Standard Ash), 10 samples / 7 min (ASTM Volatile matter 1) 10 samples / 36 min(ASTM Volatile matter 2)
|
Sample weight
|
0.5 ~ 1.1g
|
Max. power
|
4.5 kW
|
Furnace Temp
|
Room temp. to 1000℃
|
Furnace Heating
|
Resistance Wound
|
Temperature Calibration
|
Standard thermocouple
|
Temp.control precision
|
±2℃ (moisture)
±5℃ (ash, volatile matter) |
Gas Flow Rate
|
Moisture:Nitrogen or Air 2~4(Furnace Volume /min)
Volatile matter:Nitrogen 2~4(Furnace Volume /min) |
Power requirement
|
220VAC 50/60Hz
|
Gas requirement
|
Air / Nitrogen / Oxygen
|
Gas Pressure
|
Air 5 to 6 bar (75 to 90 psi)
|
Minimum Gas Purity
|
Air / Nitrogen / Oxygen 99.5%; dry, oil-free
|
Sample weighing
|
Under normal temp.
|
Test Temperature
|
Moisture 105℃,
Ash 815℃, Volatile matter 920℃~950℃ Can be customized |
Humidity
|
<80%
|
Dimensions
|
1055 x 747 x 656 mm
|
Weight
|
211kg
|
Operational Control
|
Windows-based software on external PC/Windows7/ Windows10
|
Advance Calculation
|
Yes
|
Analysis Report
|
Yes
|
Statistics Report
|
Yes
|
System Monitor
|
Yes
|