메뉴 건너뛰기

Element Analyzers

CHNS Analyzer, XRF, AES, OES, Spark-OES, AAS, AFS, ADX

SDHFCS1000 High-Frequency InfraRed Carbon & Sulfur Analyzer (고주파 적외선 탄소 및 유황 원소분석기)

본 모델은 철, 비철금속, 합금강 및 기타 금속 재료 및 광석, 토양, 배터리 재료, 핵 재료, 이산화 티타늄 및 기타 비금속 재료의 탄소 (low & high carbon) 및 황 (low & high sulfur) 함량 측정에 적합합니다. Full range에서 빠르고 정확하며 신뢰할 수있는 자동화된 측정 분석, 블라인드 샘플의 결과 자동 결정, 한 번의 분석 실행으로 high & low carbon/sulfur 의 검출 결과를 얻을 수 있습니다.

Standard: ISO 9556-1989, ISO 4935-1989, ISO 15350-2000

 

Highlights

1. High Intelligence

  • The results of high/low carbon and high/low sulfur can be obtained at the same time. Support blind sample testing, the result is automatically judged, testing efficiency improved.
  • Automatic alarm: when the channel parameters are changed, the setting conditions are wrong, the sample is not analyzed, the standard sample is not selected, the communication is abnormal, etc., the relevant alarm signal can be displayed.
  • Support networking management and data sharing.

 

2. Full Measuring Range, Easily Adaptable to Customer Requirements  

  • Adopted dual IR cells (carbon cell, low sulfur cell) to expand the measuring range of high and low carbon analysis, so that the analysis accuracy of ultra-low carbon up to ±0.2 ppm.
  • Full measuring range, from 0.00001% to 99.99%, easily adaptable to customer requirements.
  • Unique "linearization" mathematical model: non-segmented and no inflection point compensation, so that the instrument has an almost ideal linearity.
  • Automatic correction function (single point, linear): whether single point or multiple points, the data can be automatically selected to realize automatic correction in the true sense.

 

3. Large Power Combustion

  • The large power combustion furnace, using the latest generation of high-frequency component modules, overall modularization, stable output, low failure rate, easy to burning the refractory samples.
  • Double air cooling circulation system in the component area of the high-frequency furnace, low noise, high efficiency, and excellent cooling effect.
  • The surface of the high-frequency induction coil is coated with special materials, which has high induction efficiency, long life of the high-frequency oscillation tube, and the power does not decay in long-term use.

 

4. High Frequency Radiation Shielding Technology

  • The high-frequency radiation shielding function decreases harm to human body from radiation.
  • The high-frequency magnetic field has little interference to the infrared circuit board, and the test data is more accurate, especially for the measurement of ultra-low carbon and sulfur, which can effectively improve the accuracy.

 

5. Intelligent Hibernation and Self-protection

  • After the analyzer is idle for a certain time, the software automatically disconnects from the analyzer. Effectively save energy, reduce the loss of components, and extend the service life of the instrument.
  • The high-frequency furnace unit is controlled separately from the power supply of the analyzer. The analyzer can be kept power on for a long time to ensure the stability of the infrared cells. The high-frequency furnace can be stabilized after 15 minutes of preheating, which shortens the stabilization time of the instrument.

 

6. Rapid Analysis and Intelligent Dust Removal

  • The whole process of sample analysis only takes 35 seconds.
  • Integrating dust brushing, dust exhausting, and dust collection, dual-zone synchronous type and copper brush cyclone dust removal device: dust removal and sample analysis are performed simultaneously, and can be controlled manually to achieve ultra-low sulfur and high-precision analysis that is not affected by dust.
  • High pressure cyclone synchronous dust removal, with automatic dust discharge and dust accumulation function: it is cleaned after each sample analyzed to minimize the influence of dust on the results.

 

7. Stable and Reliable Gas Path, Low Operating Cost

  • Adopted high-precision gas flow controller to automatically control the gas flow without being affected by external factors such as environment, temperature and pressure.
  • Ultra-microporous metal dust filter (0.5 micron), integrally sintered and molded, can be used for a long time without ultrasonic cleaning, effectively reducing the operating cost.
  • The body is easy to disassemble and assemble, and furnace reagent is easy to replace.
  • Optional high-performance touch screen, with multi-functional retractable and foldable holder.

 

Specifications:

Description

SDHFCS1000

Measuring Range

Low carbon: 0.00001 - 0.3%

High carbon: 0.1 - 99.99%

Low sulfur:0.00001 - 0.3%

High sulfur: 0.1 - 99.99% (1or 2 IR cells per element)

Sensitivity

Carbon: 0.00001% or RSD ≤ 0.5%

Sulfur: 0.00001% or RSD ≤ 1%

Repeatability Compliance with ISO 9556 and ISO 4935
Analysis time

25~99s adjustable, normal 35s 

Sample weight 0.05 - 1.0 g
Power supply 220VAC ±5% 50Hz
Combustion temperature Max. 2000 ℃
Standard ISO 9556-1989, ISO 4935-1989, ISO 15350-2000
Dimensions 690 mm x 550 mm x 800 mm
Weight Approx. 80 kg

 

SDHFCS1000 High-Frequency InfraRed Carbon & Sulfur Analyzer (고주파 적외선 탄소 및 유황 원소분석기)


 

share
« 1 2 3 4 »
« 2 / 4 »