In order to work with onset detection below 0.02°C/min, calibration at higher sensitivity is required. This may be 0.005°C/min but calibration down to 0.003°C/min or 0.002°C/min is possible. Calibration at high sensitivity will take longer than the lower sensitivity and may often be carried out over a smaller temperature range. For example, if an onset is expected at around 100°C then a calibration range of 50-200°C may be appropriate.


전문은 첨부된 "기초자료"를 다운로드 하셔서 열람하세요.

Reference:
ARC Technical Information Sheet No. 71, Thermal Hazard Technology, UK