QCM Quartz Crystal Monitor
- Quartz crystal microbalance for thin-film measurements
- Compatible with most oscillator devices
- Mounting Flange DN40CF (2.75” O.D.), rotatable
- Linear travel 150 mm, 200 mm or 250 mm
- Bakeable up to 250°C
Introduction
This standard Quartz Crystal Microbalance can be used for in situ flux calibration in many MBE applications. Due to the linear travel it is possible to move the quartz directly under the substrate for accurate rate measurements.
The main component is a gold-plated piezo crystal which oscillates with its natural frequency. If material hits the quartz surface and sticks there the mass of the quartz will change. The change of mass results in a change of the natural frequency which will be detected.
The QCM instrument enables the measurement of thickness changes in the range of sub atom layers, i.e. < 1 nm. To overcome disturbing temperature effects the QCM is water-cooled.
Application
Typical applications for the QCM Quartz Crystal Microbalance are:
-
Flux calibration of effusion cells with regard to the cell temperature
-
Uniformity measurements using the linear travel to measure the flux rate at different positions
-
In-situ thickness control during the MBE process
Technical Data
|
Mounting flange |
DN40CF (O.D. 2.75") |
|
In vacuum lenght L |
100...500 mm, others on request |
|
Linear travel T |
150, 200, 250 mm, others on request |
|
Bakeout temperature |
up to 250 °C |
|
Oscillator connector |
BNC |
|
Cooling water |
2 x Swagelok Ø 8 mm |
|
Measurement range |
depends on oscillator device |
Dimensions
E represents the internal length of the linear travel. T represents the available linear travel.
L is defined for the situation of fully retracted linear shift as pictured in the drawing.
The length E varies with the maximum travel T. Please choose from the following table:
|
Variable T |
Variable E |
|
150 |
105 |
|
200 |
120 |
|
250 |
135 |
Specific data
For general information on CF mounting flanges see Flange and Gasketdimensions .
|
Product |
CF flange |
Linear travel |
UHV dimensions*** |
|---|---|---|---|
|
|
|
[mm] |
[mm] / [mm] |
|
QCM |
40 - |
150 - |
LxxxD35 |
|
QCM |
40 - |
200 - |
LxxxD35 |
|
QCM |
40 - |
250 - |
LxxxD35 |