메뉴 건너뛰기

전자기 물성분석기
Electromagnetic analyzers

IV1 LC Tester for Ion, Resistivity, VHR, and RDC

 

Instec designed the IV1 specifically with the industrial customers in mind. Its IV1 LC Tester.jpgelectronics are capable of exceedingly sensitive current measurement and low leakage voltage measurement. Its cables and connectors are triaxial in order to allow guarding against external leakages. The IV1 can be configured to measure any or all of the following:

 

  •  Ion Density
  •  Liquid Crystal Resistivity
  •  Voltage Holding Ratio (VHR)
  •  Residual DC (RDC)

 

Ion density and liquid crystal resistivity are critical components in the proper functioning of any LC display but especially active matrix LCDs. With too many contaminant ions or too low a resistivity, displays are prone to problems such as image retention and flicker. These measurements are useful to LC manufactures both in the development of liquid crystals and quality control. In addition, display manufacturers can use them for both incoming and in line quality inspection and as a diagnostic tool to trace down root causes of problems on the line.

 

VHR and RDC are a more direct probe for potential problems. Lowered Voltage Holding Ratios can result in noticeable flicker to a display user. Elevated Residual DC values can also produce flicker and/or image retention (or image sticking).

 

In measuring VHR, the IV1’s built in function generator outputs to the sample under test an AC pulse train of user adjustable frequency, amplitude, and duration. Simultaneously, the IV1’s internal electrometer senses changes in voltage on the sample, in particular the voltage droop between pulses. For the measurement of RDC, the IV1 applies a set DC voltage to the liquid crystal cell of user selectable magnitude and duration. The built in electrometer then records the residual DC voltage left on the sample cell, again for a user selectable duration. Since the duration of the initial DC voltage may be an hour or more, while the measurement time may be set to ten or more minutes, a single RDC determination can be a long process. With that in mind we designed the IV1’s electronics with a modular structure and options exist to add up to 8 channels.

 

Technical Specifications

Ion/Resistivity

Voltage (Min)

± 10mV

Voltage (Max)

± 10V (± 100V optional)

Resolution

100 μV

Frequency

0.01 ~ 1kHz

Current Range

100pA to 1mA

Current Resolution

1pA

Voltage Holding Ratio / Residual DC Voltage Measurement

Voltage (Min)

± 10mV

Voltage (Max)

± 10V

Resolution

100 μV

Output Current

± 100mA (Max)

Offset Voltage

± 0.1mV

Input Capacitance (Hold Mode)

< 1pF

Leak Current (Hold Mode)

< ± 0.1pA

V.G Offset (Hold Mode)

± 0.1mV

Charge Injection

± 5pC

Electrical Connections

Triaxial

Electrical Connections

Triaxial

 

 

share
« 1 2 »
« 1 / 2 »