- Monitors the output of a UV Spot curing system
- Provides indication of light guide and spot curing system performance
- Measures intensities up to 19.99 Watts/cm2
- Includes several different size light guide adaptors for maximum flexibility
- Comes in Standard and Extended intensity range versions
Uvitron provides light meters (or radiometers) designed to help users of our light curing systems measure the important parameters of the curing process. All the radiometers are NIST traceable calibrated, and are available in UVA, UVB, UVC, and Visible bands. The product pages below will help you to select the most appropriate instrument for your application.
Product Summary
As UV spot curing technology has evolved, it has become increasingly important to establish a method of measuring system performance. The EIT SpotCure UV Intensity Meter is an easy, portable, effective method of quantifying UV output.
The EIT SpotCure UV Intensity Meter is a self-contained, electro-optic instrument designed to measure and display the intensity emitted by a UV curing system. Degradation of UV lamps, light guides, and reflectors can cause decreases in intensity and create curing problems. The EIT SpotCure UV Intensity Meter is designed to provide the operator with instant feedback as to the performance of the spot curing system.
Configuration
The EIT SpotCure UV Intensity Meter is totally portable, self-contained, and battery-operated using a special lithium battery stick for extremely long life – in excess of 100,000 readings! The measurement head which contains the optics, is attached to one end of the cylindrical instrument. Light guide adapters that fit into the measurement head are available to fit most size light guides. An EIT SpotCure UV Intensity Meter can be configured to read one of the following UV transmission bandwidths: 250-260nm, 280-320nm, 320-390nm or 395-445nm.
Operation
The portable, EIT SpotCure is simple to use. Its compact, flashlight-like shape (4.60” long by 1.74” diameter) can comfortably be gripped in one hand.
In operation, the instrument is gripped in one hand while the light guide is inserted with the other hand. Once a spot curing system’s light wand is inserted into the measurement head, the “START” switch is depressed and a measurement is taken. When the “START” switch is released, the measurement is frozen and can be viewed on the LDC display. It will be held for approximately 3 minutes until the display times out or until the “START” switch is depressed again. Adapters insert into the measurement head to accommodate the standard light wands on the market. This flexibility allows the instrument to be used in a variety of monitoring applications.
Short, simple operation instructions are printed on the outside of the instrument. The EIT SpotCure UV Intensity meter is designed to withstand the rugged UV environment and extremely high intensities that can be associated with UV spot curing. It can monitor UV intensities up to 20W/cm2 for the standard version and 200W/cm2 in the Extended Range version.
Applications |
||
---|---|---|
|
Version Info |
||
---|---|---|
Part # | UV0567 | |
Version | Standard | Extended Range |
Intensity Range | UVA (20W), UVB (2W), UVC (200mW), Visible (20W) | UVA (200W), UVB (20W), Visible (200W) |
Resolution | 10mW/cm2 | 100mW/cm2 |
Specifications |
||
---|---|---|
Accuracy | +/-5% typical, +/-10% maximum | |
Display | 3 ½ digit LCD | |
Power Source | Lithium battery stick | |
Battery Life | 12,500 hours continuous operation (over 100,000 readings) | |
Overall Dimensions | 16.26cm x 4.42cm Diameter (6.40”L x 1.74” ); Measurement head – 5.14cm Diameter (2.13”) | |
Weight | 358 grams (12.8 oz.) | |
Operating Temperature | 0-70°C | |
Display Time | 2.5 to 5 minutes | |
Spectral Ranges | 250-260nm (UVC), 280-320nm (UVB), 320-390nm (UVA), or 395-445nm (Visible) | |
Case Materials | Aluminum, polyester, quartz |